UV-visible spectrophotometer UV-1800CSPC, 0.5 ,1 ,2 ,4 nm spectral bandwidth four custom installation according to user requirements, can meet the stringent requirements of pharmacopoeia |
Specifications:
Power supply: | AC 220V/50Hz,110V/60Hz |
Transmittance accuracy: | ±0.3%τ (0-100%τ) ±0.002A(0~0.5A) ±0.003A(0.5A~1A) |
Wavelength accuracy: | ±0.3nm |
Stray light: | ≤0.03%τ(220nm NaI,340nm NaNo2) |
Stability: | 0.0005/h(500nm Preheated) |
Transmittance Repeatability: | ±0.15%τ (0-100%τ) ±0.001A(0~0.5A) ±0.0015A(0.5A~1A) |
Photometric range: | -4~4A |
Power: | 120W |
Wavelength Reproducibility: | ≤0.1nm |
Wavelength range: | 190~1100nm |
Instrument Size: | 560×450×230mm |
Spectral Bandwidth: | 0.5 ,1 ,2 ,4 nm |
Display: | 6 inches high brightness LCD display |
Net weight: | 28Kg |
Detector: | Imports of silicon photodiode |
Source: | Imported deuterium lamp, tungsten lamp imports |
Wavelength tuning: | Transmittance, absorbance, concentration, energy |
Metering: | Automatic |
LSI design greatly improves the scalability and reliability of the system |
|||||||||
UV-1800CSPC series dual-beam optical system, successfully achieved the perfect combination of high precision and high reliability of the measurement can meet the requirements of various applications, can be used in biological research, biotechnology industry, pharmaceutical analysis, pharmaceutical, teaching and research, environmental protection , food hygiene, clinical testing, health and epidemic prevention and other areas | |||||||||
Improved optimization of optical design, import source and receiver creating a system of high performance and high reliability | |||||||||
UV-1800CSPC has strict requirements 0.5 ,1 ,2 ,4 nm four spectral bandwidth to meet the pharmacopoeia | |||||||||
based on user requirements optional hole rack, manual four with rack, manual Eighth rack, automatic Eighth frame, glass frame, test tube rack, 1cm colorimetric frame, 5cm colorimetric frame, 10cm colorimetric racks | |||||||||
Realize spectral scanning by PC software control, more precise and flexible measurement requirements | |||||||||
Measurement data can be output through the printer with USB interface | |||||||||
can be powered down to save the measurement parameters and data, user-friendly | |||||||||
wide wavelength range, which can meet various areas of the wavelength range of requirements | |||||||||
Rich measurement methods, with a wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination (optional), dual-wavelength, three wavelength (optional) DNA protein measurement (option) and other measuring methods, can meet different measurements requirements, and can be displayed directly on a 10-inch screen | |||||||||
Automatic design, to achieve the most simple means of measurement |